Evaluation of Consistency of Roche Electrochemiluminescence Method and Chemiluminescence Microparticle Immunoassay in Detecting Tumor Serum CA153,CA125 and CA199 |
Zhang, Fengqi |
Department of Laboratory Medicine,Sanmenxia Central Hospital,Sanmenxia 472000,China |
|
|
Abstract Objective To verify the performance ofcombined detection of three tumor markers of CA125,CA153 and CA199 by Roche Cobase602 electrochemiluminescence method in breast cancer and liver cancer patients,and to evaluate the consistency of results of electrochemiluminescence method and chemiluminescence microparticle immunoassay for detecting CA125,CA153 and CA199 in tumor diagnosis.Methods The above two chemiluminescence methods were used to detect the CA199,CA125,and CA153 of the120 tumor serum samples submitted for examination in our hospital,and the consistency of the final test results obtained by these two methods was evaluated.Results For the serum levels of CA125,CA153 and CA199 at low concentrations (0.0-35.0 U/ml),there were statistically significant differences in the results of the two methods.There were no significant differences in the detection results of the two detection methods under the condition of CA125,CA153 and CA199 at middle concentration and high concentration (35.1-500.0 U/ml,>500.0 U/ml),P>0.05.Conclusion Roche Cobase602 electrochemiluminescence method and chemiluminescencemicroparticle immunoassay have good consistency in the high-value parts of ndicators such as CA125,CA153 and CA199.Although there is no universality,it can be seen that the data among the CA125,CA153 and CA199 detection systems at low concentration are not consistent,and are consistent at middle and high concentrations.
|
Received: 29 August 2020
|
|
|
|
|
[1] 梁罕超. 应用化学发光法检测CA125和CA153诊断卵巢肿瘤的临床意义[J].临床医学,2019,39(3):40-42. [2] 刘思奇,边娜娜,苏亚娟.超声BI-RADS联合血CA15-3和CEA在乳腺肿物诊断中的价值研究[J].实用肿瘤学杂志,2020,34(4):347-351. [3] 曹春芳,邓懋清.电化学发光法检测肿瘤相关抗原结果临床意义分析[J].实用心脑肺血管病杂志,2011,19(6):992-993. [4] 朱壹澎,卢高峰,黄丽娜,等.血清CA125、CA199、CEA、AFP检测在肝炎与肝硬化鉴别诊断中的价值[J].山东医药,2019,59(1):46-48. [5] Danchen W,Songlin YU,Yicong Y,et al.Consistency evaluation of thyroid stimulating hormone results from 4 chemiluminescence assays[J].Chinese Journal of Clinical Laboratory,2019,23(2):23-26. |
[1] |
. [J]. journal1, 2021, 41(3): 63-64. |
|
|
|
|